Home > Knowledge > Content

Precautions for the use of laboratory reactors

Jan 14, 2021

 1. The laboratory reactor should be used in the designated place and operated in accordance with the instructions.

  2. Find out the test pressure, operating pressure and maximum operating temperature engraved on the main container, and use it within the allowable range of conditions.

   3. The pressure used by the pressure gauge should preferably be used within 1/2 of the indicated pressure. And often compare the pressure gauge with the standard pressure gauge for correction.

  4. The pressure gauge for oxygen should not be mixed with pressure gauges for other gases.

  5. For safety valves and other safety devices, use equipment that has undergone regular inspections and meets the specified requirements.

  6. Pay attention to the operation of the laboratory reactor, and the thermometer must be accurately inserted into the reaction solution.

  7. Keep the inside of the autoclave and the liner clean.

   8. When covering the disc flange cover, the bolts on the diagonal should be tightened in pairs.

   9. When the measuring instrument ruptures, most cases will break on the front and back sides of the glass surface. Therefore, do not stand in these dangerous places during operation. When danger is expected, remove the glass and replace it with a new one.

https://www.syhgzben.com/