1. When the laboratory reactor is out of service for a long time, the inside and outside of the reactor should be cleaned and wiped without water and other materials, and stored in a clean, dry and non-corrosive place.
2. Every time you turn on the machine, it is required that any button should be in the initial state, which requires the operator to develop good working habits, turn the knob back to the minimum position after each work is completed, to prevent the current overshoot when the next time the machine is turned on Cause major damage to the controller.
3. Regularly check various instruments and blasting relief devices to ensure their accurate and reliable work. The working environment of the equipment should meet the requirements of safety technical specifications.
4. When loading the reaction medium, it should not exceed 2/3 of the liquid level of the laboratory reactor body!
5. When reacting with different media in the reactor, first check whether the media corrodes the main material. Special orders are required for instantaneous reactions that produce a large amount of gas or high-temperature flammable and explosive chemical reactions, as well as ultra-high pressure, ultra-high temperature or medium containing chloride and fluoride ions that cause serious corrosion to stainless steel.
6. When the controller is powered on, even if the heating and stirring switches are not turned on, the output terminals on the rear panel of the controller are still live. It is not allowed to pull or touch any plugs during operation.
7. Connect the corresponding voltage according to the operating voltage of the controller, single-phase 220V, three-phase 380V, and the live wire and neutral wire of the controller should not be connected reversely; do not implement wiring work during power transmission to prevent electric shock! In order to ensure the controller For normal operation and personal safety of staff, please connect the grounding wire! The controller itself is not explosion-proof, so it should avoid flammable and explosive environments such as oil and gas.
8. When cleaning the autoclave, pay special attention not to pour water or other liquids into the heating furnace to prevent the heating furnace wire from burning. If heating the jacketed heat transfer oil, be careful not to mix water or other liquids into the heat transfer oil. Check the oil level of the heat transfer oil from time to time.
9. During or at the end of work, it is strictly forbidden to disassemble under pressure, and it is strictly forbidden to work under overpressure or overtemperature.